Home / Instruments / Torque & force

Torque wrench and force gauge calibration

Click-type torque wrenches, torque screwdrivers and torque testers, force and push-pull gauges, and load cells.

Send your instrument list and we reply with a written quotation, including the return date for each item. Collection and return within the Klang Valley.

About this calibration

  • Method

    Torque wrenches and screwdrivers are tested on a reference torque tester at three points from the bottom of their range up to their maximum, with repeat readings at each, following the approach of ISO 6789. Torque testers are checked with a reference torque transducer or a calibration arm and weights. Force gauges and load cells are tested pulling and pushing (tension and compression) against reference weights or a reference load cell. Your quotation states the test points.

  • How often

    Where you have no interval of your own, ISO 6789 recommends recalibrating a torque wrench every 12 months or 5000 cycles, whichever comes first, and after an overload, repair or drop. Your quality system sets the interval.

  • Your certificate

    Records the reading against the reference at each point, the error as a percentage and the measurement uncertainty, so you can check it against the tool's tolerance.

Instruments in this category

9 instruments and ranges
Instruments we calibrate, with range and where the work is done
InstrumentRangeWhereQuote list
Force gauge / push-pull gauge 0–10 N Laboratory
Force gauge / push-pull gauge 0–100 N Laboratory
Force gauge / push-pull gauge 0–500 N Laboratory
Load cell 0–10 kN (≈ 1000 kg) Laboratory or on site
Torque wrench, click-type 5–25 N·m Laboratory
Torque wrench, click-type 25–135 N·m Laboratory
Torque wrench, click-type 100–330 N·m Laboratory
Torque screwdriver 0.05–5 N·m Laboratory
Torque meter / tester Range confirmed on quotation Laboratory

Ranges shown are typical; exact capability is confirmed on your quotation.